Parallelism error characterization with mechanical and interferometric methods

Year: 2007

Authors: Vannoni M., Bertozzi R.

Autors Affiliation: CNR – Istituto Nazionale di Ottica Applicata, Centro SIT n. 130, Largo E. Fermi 6, 50125 Firenze, Italy;
CERMET, Centro SIT n. 52, Via Cadriano 23, 40027 Cadriano di Granarolo (BO), Italy

Abstract: Two methods for measuring the parallelism error of gauge blocks are discussed. The first method makes use of contact probes, sampling the gauge thickness at fixed locations; the parallelism error is given by the difference between maximum and minimum thickness over the sampled area. The second method is based on interferometry, producing the height maps of the opposing surfaces with respect to one another; the parallelism error is obtained after processing of the interferometric data. The two methods have been implemented in different laboratories, and compared by measuring five different gauge blocks. The results are reported, showing that the first method is more straightforward and simple, but requires prior calibration and provides information only on a limited number of sample points. The second method is more demanding in terms of equipment and measuring procedure, but in principle it is an absolute method, and provides information over a large number of data points. (C) 2006 Elsevier Ltd. All rights reserved.

Journal/Review: OPTICS AND LASERS IN ENGINEERING

Volume: 45 (6)      Pages from: 719  to: 722

KeyWords: parallelism error; gauge blocks; metrology comparison
DOI: 10.1016/j.optlaseng.2006.10.010

Citations: 8
data from “WEB OF SCIENCE” (of Thomson Reuters) are update at: 2024-11-17
References taken from IsiWeb of Knowledge: (subscribers only)
Connecting to view paper tab on IsiWeb: Click here
Connecting to view citations from IsiWeb: Click here