Hartmann test modification for measuring ophthalmic progressive lenses
Anno: 1994
Autori: Castellini C., Francini F., Tiribilli B.
Affiliazione autori: Istituto Nazionale di Ottica, Largo E. Fermi 6, Firenze 50125, Italy
Abstract: We describe a modified verion of the Hartmann test in which the pattern of the holes is replaced with a circular scanning laser beam. A position-sensitive detector is used to acquire the coordinates of the deflected beam. Significant data are obtained when the Fourier transform of the detector output signals is considered. Application of this test method to the mapping of optical parameters of ophthalmic progressive addition lenses is presented. Prismatic deviation, spherical power, and astigmatism are measured with appropriate accuracy.
Giornale/Rivista: APPLIED OPTICS
Volume: 33 (19) Da Pagina: 4120 A: 4124
Parole chiavi: Numerical analysis; Ophthalmology; Testing; Vision aids, Hartman test; Ophthalmic progressive lenses, LensesDOI: 10.1364/AO.33.004120Citazioni: 36dati da “WEB OF SCIENCE” (of Thomson Reuters) aggiornati al: 2025-05-04Riferimenti tratti da Isi Web of Knowledge: (solo abbonati) Link per visualizzare la scheda su IsiWeb: Clicca quiLink per visualizzare la citazioni su IsiWeb: Clicca qui